Commonly seen medical bag equipment include infusion bags, urine bags, blood pressure cuffs, oxygen bags, blood bags, etc. They are generally used to hold liquids or gases.
Bag products are generally made of PVC materials and processed with high-frequency heat sealing machines. If the airtightness is not good, it will cause gas or liquid leakage, which may affect the user experience at best, or threaten human life and cause immeasurable losses.

So how do we testIs there any leakage in the medical bag?Does the phenomenon exist?
Testing equipment: HC classic seriesAir tightness testing equipment

Detection method: Direct detection is adopted, and matching jigs and tooling are customized according to customer needs.Place the product to be tested on the fixture and press the instrument start button to start testing.

Testing parameters: inflation 15Kpa, time 10s, voltage stabilization 10s, test 5s.During this process, if the leakage exceeds the upper limit of 50pa, the program triggers an alarm or an NG signal appears, that is, the air tightness of the product to be tested fails, and it is a defective product.
Reasons to choose Hai Ruisi
Hai RuisiAir tightness testing equipmentAdvantages:
The mold processing accuracy is high, up to 0.02mm;
Air tightness detector with resolution up to 0.1pa;
Hairisi has all product lines of air tightness detectors, true water detectors, and leakage point detectors, which can meet all sealing testing usage scenarios from trial production to mass production;
Hairuisi has the ability to quickly integrate systems and can deliver the goods as soon as 7 days after placing the order;
The designed detection system strictly considers the stability and accuracy of the test results.
Advantages of Hairis air tightness testing equipment:
High precision ensures the stability of the mold;
High resolution, even the smallest leaks can be accurately detected;
All requirements for sealing testing can be packaged and purchased at Hairis, reducing the total cost of ownership;
Hairis can respond quickly. When your internal process is completed, the testing equipment has just been produced;
Strict requirements bring a stable detection system to help you select 100% defective products.