A medical device endoscope is a tube equipped with light that enters the human body through a natural orifice or a small surgical incision; some endoscopes even have micro-integrated circuit sensors that feed the observed information back to the computer.

In order to ensure the normal use of the endoscope, the air tightness test process must be carried out to strictly control the sealing performance of the endoscope.
Today I’m going to talk aboutEndoscope air tightness testingcase.
Testing instrument: HC classic series air tightness tester

Detection method: Customize fixtures and tooling according to the structure of the endoscope and customer needs.

Using the testing method of finished product and large leakage volume determination, put the product to be tested into the mold, press the start button of the tooling, and start testing.Inflation 200Kpa, inflation time 8s, pressure stabilization time 10s, test time 5s, leakage upper limit 30pa.During the test, if the leakage value exceeds the set value, the product to be tested is a defective product, otherwise it is a qualified product.
Hai RuisiAir tightness testing equipmentAdvantages:
1. The patented technology "IVT pneumatic sealing valve island" independently developed by Hai Ruisi greatly improves accuracy.
2. Advanced differential pressure software algorithm, the minimum resolution can reach 0.1Pa.
3. It has 100 groups of subroutines, and each group of subroutines can be programmed independently.
4. Full-touch large-screen intelligent operating system with simple and clear operation interface.
5. The standard volume recognition function can automatically identify the volume of the sample to be measured.
6. Optional quick inflation function.

Hai RuisiSealing testing equipmentincome:
A. Ensure the stability of the mold;
B. High resolution, no matter how small the leakage is, it can be accurately detected;
C. Experienced engineers bring sophisticated mold design, so that every set of molds you use is stable and reliable;
D. Instant communication with Haireisi can save you the pain of looking for contacts everywhere;
E. Strict requirements bring a stable detection system to help you select defective products.